Autoclave Glossary
43 terms you will meet in autoclave quotations, brochures and spec sheets — defined in plain English.
A–Z of terms
- A0 value
- The measure of thermal disinfection in a washer-disinfector: equivalent seconds at 80 °C. 90 °C held for 5 minutes gives A0 3000, the usual target for instruments. Full guide
- Air removal
- Getting air out of the chamber and load so steam can touch every surface — by vacuum pulses or steam displacement. Full guide
- Autoclave
- A steriliser that uses saturated steam under pressure to kill all microorganisms, including spores. Full guide
- Batch control identification
- Recording the steriliser, cycle number and date of a pack in the patient’s notes, so instruments can be traced to patients if a cycle later fails. Full guide
- Biological indicator
- A test containing resistant bacterial spores (Geobacillus stearothermophilus) that shows whether a cycle actually killed them. Full guide
- Bowie-Dick test
- A daily test pack for vacuum sterilisers that shows whether air removal and steam penetration are working. Full guide
- Chemical indicator
- A strip or ink that changes colour when exposed to sterilising conditions; types 1–6 under ISO 11140-1 differ in how much they prove. Full guide
- Class B
- EN 13060 class for small sterilisers proven on wrapped, porous and hollow loads, using fractionated pre-vacuum. Full guide
- Class N
- EN 13060 class proven only for unwrapped solid instruments, using passive air removal. Full guide
- Class S
- EN 13060 class for the specific load types the manufacturer has validated. Full guide
- Cold sterilisation
- An old name for soaking instruments in a liquid chemical such as glutaraldehyde; most soaks achieve high-level disinfection, not sterilisation. Full guide
- Conductivity
- A measure of dissolved ions in water, in µS/cm; autoclave feed water is commonly specified at around 15 µS/cm or lower. Full guide
- Door seal (gasket)
- The silicone ring that seals the chamber door; dirt or wear on it is the most common cause of failed vacuum and leak tests. Full guide
- Double-door (pass-through) steriliser
- A large steriliser built into a wall, loaded from the packing room and unloaded into the sterile store, with interlocked doors. Full guide
- Dry heat sterilisation
- Sterilising with hot air, typically 160 °C for 2 hours to 180 °C for 30 minutes — for glass, oils, powders and items steam would corrode. Full guide
- Drying phase
- The end of the cycle that removes moisture from the load; wet packs are not treated as sterile. Full guide
- EN 13060
- European standard for small steam sterilisers with chambers up to 60 litres. Full guide
- EN 285
- European standard for large steam sterilisers that can hold at least one 300 × 300 × 600 mm sterilisation module — the hospital CSSD class. Full guide
- Establishment licence
- The MDA licence that manufacturers, importers, distributors and authorised representatives of medical devices need under Act 737. Full guide
- Ethylene oxide (EO)
- A toxic gas used for low-temperature sterilisation of heat-sensitive devices; loads need hours of aeration afterwards. Full guide
- Foil test
- A quick ultrasonic cleaner check: aluminium foil held in the running tank should come out evenly pitted; untouched patches show dead zones. Full guide
- Fractionated vacuum
- Several alternating vacuum and steam pulses that pull air out of hollow and porous items. Full guide
- Heat sealer
- A machine that closes heat-seal pouches and reels at a set temperature, pressure and time; the seal it makes is part of the sterile barrier. Full guide
- Helix test
- A hollow-load test device (process challenge device) that checks steam reaches the end of a long narrow tube. Full guide
- Hold time
- The time the load is kept at sterilising temperature — at least 3 minutes at 134 °C or 15 minutes at 121 °C. Full guide
- Hydrogen peroxide gas plasma
- A low-temperature sterilisation method using peroxide vapour and plasma, for heat-sensitive devices; it cannot process cellulose or liquids. Full guide
- ISO 17665
- International standard for developing, validating and routinely controlling moist-heat sterilisation. Full guide
- Latent heat
- The heat steam releases when it condenses back into water — about 2,200 kJ per kilogram — which is why steam heats instruments far faster than hot air. Full guide
- Liquids cycle
- A steriliser programme for bottled liquids and media that skips vacuum pulses and releases pressure slowly (slow exhaust), so superheated liquid does not boil over. Full guide
- Prion cycle
- An extended 134 °C cycle, typically 18 minutes, used where infection-control policy requires it. Full guide
- Process challenge device (PCD)
- A test pack or device that holds an indicator behind a defined resistance to steam, so a routine test is at least as hard as the hardest item in a real load. Full guide
- Saturated steam
- Steam at the exact temperature and pressure where it carries maximum heat; wet or superheated steam sterilises poorly. Full guide
- Self-contained biological indicator
- A vial holding spores and a sealed ampoule of growth medium; after the cycle the ampoule is crushed and the vial incubated, and a colour change shows surviving spores. Full guide
- Small-value asset
- An asset cheap enough to be written off for capital allowance in the year it is bought — up to RM2,000 each now, RM3,000 proposed in Budget 2027. Full guide
- Spaulding classification
- Sorts items into critical, semi-critical and non-critical to decide whether they need sterilising or disinfecting. Full guide
- Standard operating procedure (SOP)
- The written, approved steps staff follow to run, test and record the steriliser, and to act on failures. Full guide
- Sterilisation pouch
- A paper-film pack that lets steam in and keeps the instrument sterile in storage until opened. Full guide
- SW404
- Malaysia’s scheduled-waste code for pathogenic and clinical wastes and quarantined materials, collected and treated only by DOE-licensed contractors. Full guide
- Ultrasonic cleaner
- A tank that cleans instruments by cavitation, reaching hinges and serrations that brushes miss — a cleaning step, not a steriliser. Full guide
- Validation (IQ/OQ/PQ)
- Installation, operational and performance qualification proving a specific steriliser sterilises specific loads. Full guide
- Washer-disinfector
- A machine that cleans and thermally disinfects instruments before packing — cleaning always comes before sterilising. Full guide
- Wet pack
- A pack with visible moisture after the cycle; it is not treated as sterile and must be reprocessed. Full guide
Start with these guides
Class B Autoclave vs Class N and Class SThe EN 13060 classes decide which loads your autoclave can legitimately sterilise. Here is what each class covers and how to tell them apart.Dental Autoclave for Malaysian Dental ClinicsWhy dental clinics need a Class B autoclave, how to size it by chairs and patients, and the daily routine that keeps handpieces safe.Autoclave Price in Malaysia: What Drives the CostWe do not publish a price list, because the right number depends on your configuration. Here is everything that moves the price, so you can compare quotes fairly.
Get a written quotation
Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.