Autoclave Tape, Sterilization Pouches and Indicators
What each autoclave consumable proves, how to choose and pack sterilisation pouches, and how to store them so instruments stay sterile.
What each consumable proves
ISO 11140-1 sorts chemical indicators into six types. The higher the type, the more of the cycle it responds to.
| Item | ISO 11140-1 type | What it shows | What it does not show |
|---|---|---|---|
| Autoclave tape | Type 1 (process) | The item was exposed to a cycle | That the contents are sterile |
| Printed pouch indicator | Type 1 (process) | The pouch went through a cycle | That steam reached the instrument |
| Bowie-Dick test sheet or pack | Type 2 (specific test) | Air removal and steam penetration in a vacuum unit | Anything about patient loads |
| Multi-variable indicator | Type 4 | Two or more critical variables were reached | Full cycle performance |
| Integrator strip | Type 5 | All critical variables, set against biological performance | It does not replace spore tests |
| Emulating indicator | Type 6 | The parameters of a specific stated cycle were met | Performance of other cycles |
Choosing sterilisation pouches
Self-seal or heat-seal
Self-seal pouches have an adhesive strip and need no equipment. Heat-seal pouches or reels need a sealer but let you cut any length.
Made for steam
Use pouches specified for steam sterilisation, ideally with a printed steam indicator.
Right size
The instrument should fit with space around it and without stretching the seal. Sharp tips can pierce the pouch — use tip protectors where needed.
Read next Heat sealers: rotary vs impulse
Packing, labelling and storage
- Clean and dry instruments before pouching; open hinged instruments.
- Do not overfill — steam must reach every surface.
- Write the date, cycle number and contents outside the sealed area.
- Let pouches dry fully in the autoclave; reprocess any that come out wet or torn.
- Store in closed, clean, dry cupboards off the floor and away from sinks.
Read next Bowie-Dick, spore tests and validationCara guna autoclave (Bahasa Malaysia)How long pouched instruments stay sterileSpore test incubators and BI kits
Related guides
Autoclave supply near you
Written quotes for clinics across Malaysia, handled from Batu Caves, Kota Kinabalu and Kuching.
Frequently asked questions
Can I fold a pouch and close it with autoclave tape?
Do I still need integrators if my pouches have indicators?
Can sterilisation pouches be reused?
Get a written quotation
Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.
010-905 2891 · Mon–Sat, 9:00 am – 6:00 pm