Autoclave Specifications: How to Read a Spec Sheet and Write One for a Tender

Two autoclaves with the same litre figure can be very different machines. What each line of a spec sheet tells you, which numbers to check against your room, and a specification table you can adapt for a quotation request or tender.

Also called Autoclave specification, Spec sheet, Tender specification, Technical specification

Published:

Start from the type-test table, not the litres

EN 13060 sorts small steam sterilisers into Type B, N and S cycles, and a good spec sheet shows which type tests each programme has passed. MELAG’s Vacuklav 24 BL+ manual, for example, lists five programmes and ticks the loads each passed — solid, porous, narrow-lumen hollow (hollow load A), simple hollow (hollow load B), single and multiple wrapping, and drying.

  • Check that the programme you will use every day has passed the tests for your loads. On that unit, the Type S quick programme is not tested for porous, wrapped or narrow-lumen loads, and even the Type B quick programme is not tested for porous loads; the Universal programme covers all three.
  • Read the temperature and holding time of each programme — for example 134 °C for 5½ minutes against 121 °C for 20½ minutes on the same unit — and ask for the total cycle time including drying.
  • If a sheet only says “Class B” with no type-test table, ask for it in writing.

The rest of the spec sheet, line by line

Examples are from the MELAG Vacuklav 24 BL+ technical data, to show what a complete sheet contains. Other models will differ.

Line on the sheetExampleCheck it against
Chamber size and volume25 cm diameter × 60 cm deep, 28.6 litresHow many of your trays or cassettes actually fit — chamber volume is not usable space
Electrical supply220–240 V, 50/60 Hz, 2,100 W; 16 A building fuse and 30 mA residual-current protectionYour socket and circuit; Malaysian single-phase supply is 230 V, 50 Hz
Feed waterDemineralised water from a reservoir or water treatment unit, with built-in conductivity monitoringYour water treatment and the maker’s purity limit
Other connectionsA cooling-water connection with drinking water at a stated flow and pressurePlumbing and drainage at the bench — not every model needs them
Size and weight49.5 × 42.5 × 83.5 cm (height × width × depth); 53 kg empty, 60 kg in useBench depth and strength, plus the clearances in the manual
Room conditions5–40 °C; humidity limits; waste heat; 63 dB(A) at 1 mVentilation and air-conditioning in a Malaysian room
RecordsPrinter, data storage or networkYour logbook and traceability system
Built-in testsA vacuum test programme; the steam-penetration test is run in the Universal programme with a test deviceThe daily and weekly tests your staff will run

Power and water in a Malaysian clinic

Power

TNB supplies single-phase premises at 230 V, 50 Hz. A 13 A socket carries at most about 3 kW at 230 V, so check the unit’s rated power and the maker’s circuit requirements before delivery day. Some makers ask for a dedicated circuit and residual-current protection.

Water

Small sterilisers run on demineralised or distilled water within the maker’s purity limit, and some need a drain or a cooling-water supply. Plan the water treatment and plumbing before the unit arrives.

Heat and space

A steriliser gives off heat and needs clearance around its vents. Check the ambient limits on the sheet against the room, especially if the air-conditioning is switched off overnight.

A specification you can adapt

Write requirements as outcomes and minimums, not one brand’s numbers, so more than one registered model can comply. Replace the brackets with your own figures.

ItemExample requirement
TypeTabletop steam steriliser with fractionated pre-vacuum, meeting EN 13060 for Type B cycles
Programmes134 °C and 121 °C programmes; type-test results provided for hollow load A, porous and wrapped loads for the programme used
CapacityUsable space for at least [n] standard trays or [n] cassettes of [size]
Cycle timeTotal time including drying stated for a full wrapped load
TestsBuilt-in vacuum leak test and Bowie-Dick or helix programme
WaterAutomatic feed from a reservoir or demineraliser, with feed-water quality monitoring; maker’s water limits stated
RecordsRecord of every cycle with batch number, date, time, temperature and pressure; printout or export by USB or network
PowerSingle-phase 230 V, 50 Hz; rated power and circuit protection stated
Installation spaceFits a bench [width × depth] cm including the maker’s clearances; operating weight stated
RegistrationMDA registration number for the model, and the licensed establishment supplying it
Installation and trainingInstallation, IQ and OQ records, first tests at handover and user training, with who performs each
ServiceWarranty period, preventive maintenance interval, breakdown response time, parts held in Malaysia, loan unit
DOSHSupplier to state whether the model needs DOSH design verification or a certificate of fitness, and who applies

Documents to ask for

Autoclaves and DOSH

The Occupational Safety and Health (Plant Requiring Certificate of Fitness) Regulations 2024, in operation since 1 June 2024, mention autoclaves by name. The documents for design verification include an autoclave’s piping and instrumentation diagram, its name plate and, if imported, its catalogue; verification by an authorised inspecting body is not required for an autoclave’s design drawings and calculations; and the rule that a steam boiler must be run by a certified boiler engineer and operator does not apply to autoclaves. Some low-pressure vessels are excluded altogether, so whether a model needs design verification or a certificate of fitness depends on the model. Get the supplier’s answer in writing.

  • The MDA registration number, and a dated screenshot of the model on the MMDR.
  • The establishment licence and GDPMD certificate of the company supplying it.
  • The declaration of conformity and the EN 13060 type-test summary.
  • The instructions for use and the full technical data sheet, in English or Bahasa Malaysia.
  • An installation and qualification plan, and a sample service report.
  • A written statement on DOSH requirements for the model.

Common specification mistakes

  • Specifying litres only, instead of the trays or cassettes the chamber must hold.
  • Copying one brand’s dimensions or features, so only that model can comply.
  • Asking for “Class B” without naming the loads and asking for type-test evidence.
  • Leaving out records, which you need to release loads and trace instruments.
  • Forgetting water treatment, drainage and the electrical circuit.
  • No service terms: response time, local parts and a loan unit.
  • No MDA or DOSH documents requested until after the order.

Frequently asked questions

What specifications should I look for in an autoclave?
The EN 13060 type-test results for your loads, usable chamber space, total cycle time including drying, water and electrical requirements, size and clearances, how cycles are recorded, built-in tests, MDA registration and the service terms.
What is the difference between chamber volume and usable volume?
Chamber volume is the whole cylinder; usable space is what your trays or cassettes actually occupy once the rack is in. Specify the number and size of trays, not litres.
Does an autoclave need DOSH approval in Malaysia?
The 2024 certificate-of-fitness regulations name autoclaves in their design-verification requirements, while excluding some low-pressure vessels. Whether your model needs design verification or a certificate of fitness depends on the model, so ask the supplier to state it in writing.
Can I plug an autoclave into a normal 13 A socket?
Only if its rated power and the maker’s circuit requirements allow it. A 13 A socket carries at most about 3 kW at 230 V, and some makers ask for a dedicated circuit with residual-current protection.
How do I write an autoclave tender specification without naming a brand?
State the loads, the minimum capacity in trays, the cycle types and tests, water and power limits, records, registration and service requirements, and ask for evidence. Avoid one model’s exact dimensions and feature names.

Get a written quotation

Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.