Autoclave Specifications: How to Read a Spec Sheet and Write One for a Tender
Two autoclaves with the same litre figure can be very different machines. What each line of a spec sheet tells you, which numbers to check against your room, and a specification table you can adapt for a quotation request or tender.
Start from the type-test table, not the litres
EN 13060 sorts small steam sterilisers into Type B, N and S cycles, and a good spec sheet shows which type tests each programme has passed. MELAG’s Vacuklav 24 BL+ manual, for example, lists five programmes and ticks the loads each passed — solid, porous, narrow-lumen hollow (hollow load A), simple hollow (hollow load B), single and multiple wrapping, and drying.
- Check that the programme you will use every day has passed the tests for your loads. On that unit, the Type S quick programme is not tested for porous, wrapped or narrow-lumen loads, and even the Type B quick programme is not tested for porous loads; the Universal programme covers all three.
- Read the temperature and holding time of each programme — for example 134 °C for 5½ minutes against 121 °C for 20½ minutes on the same unit — and ask for the total cycle time including drying.
- If a sheet only says “Class B” with no type-test table, ask for it in writing.
Read next Class B autoclaves explainedPre-vacuum or gravity cycles
The rest of the spec sheet, line by line
Examples are from the MELAG Vacuklav 24 BL+ technical data, to show what a complete sheet contains. Other models will differ.
| Line on the sheet | Example | Check it against |
|---|---|---|
| Chamber size and volume | 25 cm diameter × 60 cm deep, 28.6 litres | How many of your trays or cassettes actually fit — chamber volume is not usable space |
| Electrical supply | 220–240 V, 50/60 Hz, 2,100 W; 16 A building fuse and 30 mA residual-current protection | Your socket and circuit; Malaysian single-phase supply is 230 V, 50 Hz |
| Feed water | Demineralised water from a reservoir or water treatment unit, with built-in conductivity monitoring | Your water treatment and the maker’s purity limit |
| Other connections | A cooling-water connection with drinking water at a stated flow and pressure | Plumbing and drainage at the bench — not every model needs them |
| Size and weight | 49.5 × 42.5 × 83.5 cm (height × width × depth); 53 kg empty, 60 kg in use | Bench depth and strength, plus the clearances in the manual |
| Room conditions | 5–40 °C; humidity limits; waste heat; 63 dB(A) at 1 m | Ventilation and air-conditioning in a Malaysian room |
| Records | Printer, data storage or network | Your logbook and traceability system |
| Built-in tests | A vacuum test programme; the steam-penetration test is run in the Universal programme with a test device | The daily and weekly tests your staff will run |
Read next Setting up the sterilisation roomWater quality for autoclaves
Power and water in a Malaysian clinic
Power
TNB supplies single-phase premises at 230 V, 50 Hz. A 13 A socket carries at most about 3 kW at 230 V, so check the unit’s rated power and the maker’s circuit requirements before delivery day. Some makers ask for a dedicated circuit and residual-current protection.
Water
Small sterilisers run on demineralised or distilled water within the maker’s purity limit, and some need a drain or a cooling-water supply. Plan the water treatment and plumbing before the unit arrives.
Heat and space
A steriliser gives off heat and needs clearance around its vents. Check the ambient limits on the sheet against the room, especially if the air-conditioning is switched off overnight.
A specification you can adapt
Write requirements as outcomes and minimums, not one brand’s numbers, so more than one registered model can comply. Replace the brackets with your own figures.
| Item | Example requirement |
|---|---|
| Type | Tabletop steam steriliser with fractionated pre-vacuum, meeting EN 13060 for Type B cycles |
| Programmes | 134 °C and 121 °C programmes; type-test results provided for hollow load A, porous and wrapped loads for the programme used |
| Capacity | Usable space for at least [n] standard trays or [n] cassettes of [size] |
| Cycle time | Total time including drying stated for a full wrapped load |
| Tests | Built-in vacuum leak test and Bowie-Dick or helix programme |
| Water | Automatic feed from a reservoir or demineraliser, with feed-water quality monitoring; maker’s water limits stated |
| Records | Record of every cycle with batch number, date, time, temperature and pressure; printout or export by USB or network |
| Power | Single-phase 230 V, 50 Hz; rated power and circuit protection stated |
| Installation space | Fits a bench [width × depth] cm including the maker’s clearances; operating weight stated |
| Registration | MDA registration number for the model, and the licensed establishment supplying it |
| Installation and training | Installation, IQ and OQ records, first tests at handover and user training, with who performs each |
| Service | Warranty period, preventive maintenance interval, breakdown response time, parts held in Malaysia, loan unit |
| DOSH | Supplier to state whether the model needs DOSH design verification or a certificate of fitness, and who applies |
Documents to ask for
Autoclaves and DOSH
The Occupational Safety and Health (Plant Requiring Certificate of Fitness) Regulations 2024, in operation since 1 June 2024, mention autoclaves by name. The documents for design verification include an autoclave’s piping and instrumentation diagram, its name plate and, if imported, its catalogue; verification by an authorised inspecting body is not required for an autoclave’s design drawings and calculations; and the rule that a steam boiler must be run by a certified boiler engineer and operator does not apply to autoclaves. Some low-pressure vessels are excluded altogether, so whether a model needs design verification or a certificate of fitness depends on the model. Get the supplier’s answer in writing.
- The MDA registration number, and a dated screenshot of the model on the MMDR.
- The establishment licence and GDPMD certificate of the company supplying it.
- The declaration of conformity and the EN 13060 type-test summary.
- The instructions for use and the full technical data sheet, in English or Bahasa Malaysia.
- An installation and qualification plan, and a sample service report.
- A written statement on DOSH requirements for the model.
Read next How to check MDA registrationVetting the supplierIQ, OQ and PQ
Common specification mistakes
- Specifying litres only, instead of the trays or cassettes the chamber must hold.
- Copying one brand’s dimensions or features, so only that model can comply.
- Asking for “Class B” without naming the loads and asking for type-test evidence.
- Leaving out records, which you need to release loads and trace instruments.
- Forgetting water treatment, drainage and the electrical circuit.
- No service terms: response time, local parts and a loan unit.
- No MDA or DOSH documents requested until after the order.
Read next How buying an autoclave worksBrands you will see in MalaysiaSpesifikasi autoclave (Bahasa Malaysia)
Related guides
Autoclave supply near you
Written quotes for clinics across Malaysia, handled from Batu Caves, Kota Kinabalu and Kuching.
Sources and further reading
- W&H — Types of sterilizers: EN 13060 Types B, S and N, EN 285 and ANSI/AAMI ST55
- MELAG — User Manual, Vacuklav 24 BL+ steam sterilizer (servicing intervals, spare parts and technical data)
- Occupational Safety and Health (Plant Requiring Certificate of Fitness) Regulations 2024, P.U.(A) 99/2024 — in operation 1 June 2024
- DOSH — Application Guide for Design Verification of Imported Pressure Vessel
- Tenaga Nasional Berhad — Electricity Supply Application Handbook: metering guidelines (230 V single-phase supply)
- Medical Device Authority — Malaysia Medical Device Register (MMDR), public search
Frequently asked questions
What specifications should I look for in an autoclave?
What is the difference between chamber volume and usable volume?
Does an autoclave need DOSH approval in Malaysia?
Can I plug an autoclave into a normal 13 A socket?
How do I write an autoclave tender specification without naming a brand?