Autoclave Temperature and Pressure: 121 °C vs 134 °C
What the numbers on an autoclave display mean, why steam pressure and temperature move together, and which time at which temperature actually sterilises.
Last updated:
Temperature and pressure are linked
A working autoclave fills its chamber with saturated steam — steam in balance with water at its boiling point. For saturated steam, every temperature has exactly one matching pressure: raise the pressure and water boils hotter, which is why an autoclave is a pressure vessel. If the temperature and pressure on a printout do not match the steam table, the chamber holds air or superheated steam and the load is at risk.
| Sterilising temperature | Absolute pressure | Gauge pressure | Gauge, in psi |
|---|---|---|---|
| 121 °C | ≈ 2.05 bar (205 kPa) | ≈ 1.04 bar (104 kPa) | ≈ 15 psi |
| 126 °C | ≈ 2.39 bar (239 kPa) | ≈ 1.38 bar (138 kPa) | ≈ 20 psi |
| 134 °C | ≈ 3.04 bar (304 kPa) | ≈ 2.03 bar (203 kPa) | ≈ 29–30 psi |
Gauge vs absolute pressure
Gauge pressure
Measured above the surrounding air, so a reading of zero means normal atmospheric pressure. Most clinic autoclave dials and printouts use gauge pressure, which is why a 134 °C cycle shows roughly 2 bar.
Absolute pressure
Measured from a perfect vacuum. Steam tables are written this way, and many Class B printouts show their vacuum pulses in absolute terms because a vacuum is simply a pressure below atmospheric.
Units on printouts
1 bar = 100 kPa = 0.1 MPa ≈ 14.5 psi. Different brands print bar, kPa or MPa — check whether a figure is gauge or absolute before comparing two machines.
Hold time: time at temperature, not cycle length
The hold (plateau) starts only when the chamber — and in a well-designed cycle, the load — is inside the sterilising temperature band. Air removal, heat-up, exhaust and drying come on top, which is why a 3-minute hold sits inside a 30- to 60-minute cycle. If the temperature drops below the band during the hold, the cycle should fail rather than quietly stretch.
| Temperature band | Minimum hold | Where you will see it |
|---|---|---|
| 121–124 °C | 15 minutes | Autoclavable plastics, rubber and textiles; lab liquids |
| 126–129 °C | 10 minutes | Some hospital and older sterilisers |
| 134–137 °C | 3 minutes | The standard instrument cycle on clinic tabletop units |
| 134 °C extended | 18 minutes | Prion cycle, where infection-control policy requires it |
| 132 °C | 4 minutes | Prevacuum cycles in US-published instructions |
Read next How steam sterilisation worksTests that prove the cycle worked
Liquids and heavy loads lag behind the chamber
- A bottle of liquid heats far more slowly than the chamber air, so the display can read 121 °C long before the centre of a large bottle gets there.
- Liquid cycles therefore use longer holds chosen by the largest container, or a load probe in a reference bottle that starts the hold only when the liquid itself is at temperature.
- Exhaust must be slow so superheated liquid does not boil over, and the door must stay shut until the liquid has cooled.
- Clinic tabletop units without a validated liquids programme should not be used for liquids.
- Dense instrument trays and heavy packs lag too — stay within the maker’s maximum load per tray.
Read next Laboratory autoclaves and liquid cycles
Material limits — the instructions for use decide
Under ISO 17664, device makers must publish reprocessing instructions, including the sterilisation temperature and time their product tolerates. The instrument’s instructions for use (IFU) — not the autoclave — set the limit.
Metal instruments
Stainless-steel instruments generally tolerate 134 °C. Carbon-steel and plated instruments can pit or corrode, so follow their IFU.
Plastics and rubber
Polypropylene and silicone tolerate 121 °C and often 134 °C; many other plastics soften or warp. If the IFU says 121 °C, never run 134 °C to save time.
Handpieces and powered devices
Dental handpieces and motors carry a maximum temperature in their IFU. Exceeding it shortens bearing and seal life.
Heat-sensitive devices
Devices that cannot take 121 °C need a validated low-temperature method or single-use replacements.
Plasma and ethylene oxide sterilisersRelated guides
Autoclave supply near you
Delivery, installation and training across Malaysia from Batu Caves, Kota Kinabalu and Kuching.
Frequently asked questions
What temperature does an autoclave reach?
What is autoclave pressure in psi?
Is 134 °C better than 121 °C?
Why can an autoclave show the right temperature and still fail the Bowie-Dick test?