Autoclave Temperature and Pressure: 121 °C vs 134 °C

What the numbers on an autoclave display mean, why steam pressure and temperature move together, and which time at which temperature actually sterilises.

Also called Autoclave temperature, Autoclave pressure, Suhu autoclave

Published:

Last updated:

Temperature and pressure are linked

A working autoclave fills its chamber with saturated steam — steam in balance with water at its boiling point. For saturated steam, every temperature has exactly one matching pressure: raise the pressure and water boils hotter, which is why an autoclave is a pressure vessel. If the temperature and pressure on a printout do not match the steam table, the chamber holds air or superheated steam and the load is at risk.

Saturated steam near sea level. Gauge pressure = absolute pressure minus about 1.01 bar of atmosphere.
Sterilising temperatureAbsolute pressureGauge pressureGauge, in psi
121 °C≈ 2.05 bar (205 kPa)≈ 1.04 bar (104 kPa)≈ 15 psi
126 °C≈ 2.39 bar (239 kPa)≈ 1.38 bar (138 kPa)≈ 20 psi
134 °C≈ 3.04 bar (304 kPa)≈ 2.03 bar (203 kPa)≈ 29–30 psi

Gauge vs absolute pressure

Gauge pressure

Measured above the surrounding air, so a reading of zero means normal atmospheric pressure. Most clinic autoclave dials and printouts use gauge pressure, which is why a 134 °C cycle shows roughly 2 bar.

Absolute pressure

Measured from a perfect vacuum. Steam tables are written this way, and many Class B printouts show their vacuum pulses in absolute terms because a vacuum is simply a pressure below atmospheric.

Units on printouts

1 bar = 100 kPa = 0.1 MPa ≈ 14.5 psi. Different brands print bar, kPa or MPa — check whether a figure is gauge or absolute before comparing two machines.

Hold time: time at temperature, not cycle length

The hold (plateau) starts only when the chamber — and in a well-designed cycle, the load — is inside the sterilising temperature band. Air removal, heat-up, exhaust and drying come on top, which is why a 3-minute hold sits inside a 30- to 60-minute cycle. If the temperature drops below the band during the hold, the cycle should fail rather than quietly stretch.

Temperature bandMinimum holdWhere you will see it
121–124 °C15 minutesAutoclavable plastics, rubber and textiles; lab liquids
126–129 °C10 minutesSome hospital and older sterilisers
134–137 °C3 minutesThe standard instrument cycle on clinic tabletop units
134 °C extended18 minutesPrion cycle, where infection-control policy requires it
132 °C4 minutesPrevacuum cycles in US-published instructions

Choosing a steriliser? Tell us what you sterilise and we match a Class B or Class N model with 134 °C and 121 °C programs, and send the price in writing.

Liquids and heavy loads lag behind the chamber

  • A bottle of liquid heats far more slowly than the chamber air, so the display can read 121 °C long before the centre of a large bottle gets there.
  • Liquid cycles therefore use longer holds chosen by the largest container, or a load probe in a reference bottle that starts the hold only when the liquid itself is at temperature.
  • Exhaust must be slow so superheated liquid does not boil over, and the door must stay shut until the liquid has cooled.
  • Clinic tabletop units without a validated liquids programme should not be used for liquids.
  • Dense instrument trays and heavy packs lag too — stay within the maker’s maximum load per tray.

Material limits — the instructions for use decide

Under ISO 17664, device makers must publish reprocessing instructions, including the sterilisation temperature and time their product tolerates. The instrument’s instructions for use (IFU) — not the autoclave — set the limit.

Metal instruments

Stainless-steel instruments generally tolerate 134 °C. Carbon-steel and plated instruments can pit or corrode, so follow their IFU.

Plastics and rubber

Polypropylene and silicone tolerate 121 °C and often 134 °C; many other plastics soften or warp. If the IFU says 121 °C, never run 134 °C to save time.

Handpieces and powered devices

Dental handpieces and motors carry a maximum temperature in their IFU. Exceeding it shortens bearing and seal life.

Heat-sensitive devices

Devices that cannot take 121 °C need a validated low-temperature method or single-use replacements.

Plasma and ethylene oxide sterilisers

Frequently asked questions

What temperature does an autoclave reach?
Clinic autoclaves sterilise at 121 °C or 134 °C, with 134 °C the standard for most metal instruments. The hold is at least 15 minutes at 121 °C or 3 minutes at 134 °C.
What is autoclave pressure in psi?
For saturated steam, about 15 psi above atmospheric at 121 °C and about 29–30 psi above atmospheric at 134 °C.
Is 134 °C better than 121 °C?
It is faster, not better. Both sterilise when held for the right time; 134 °C for 3 minutes suits heat-tolerant instruments, while 121 °C for 15 minutes is gentler on plastics, rubber and textiles.
Why can an autoclave show the right temperature and still fail the Bowie-Dick test?
The sensor reads the chamber or drain, not the inside of a pack. Trapped air keeps parts of the load cooler even when the display looks right — exactly what the air-removal test is designed to catch.