Spore Test Incubators and Biological Indicators: What to Buy for Your Autoclave

Weekly spore testing needs three things that must work together: the indicator, the incubator or reader, and the record. How the options differ, where brand lock-in hides, and how to compare the real cost per test.

Also called Spore test incubator, Biological indicator reader, Autoclave spore test kit

Published:

What a spore-testing set-up is made of

This page is about choosing the kit. For how to place, run and read a spore test, see the method guide linked below.

Biological indicator (BI)

A known number of highly resistant spores on a carrier in a protective package. If the spores fail to grow after the cycle, the cycle was lethal enough. Steam BIs use Geobacillus stearothermophilus.

Control BI

An unprocessed indicator from the same lot, incubated alongside the test. It must grow, which proves the spores were alive and the medium and incubator work. CDC’s dental guidance calls for a matching control with each test.

Incubator or auto-reader

Holds the indicator at the temperature its instructions specify. A simple dry-block incubator is read by eye; an auto-reader detects growth itself and can store or send the result.

Activator and record

Most self-contained BIs are crushed or activated after the cycle to release their growth medium. The result belongs in your steriliser logbook with the lot number and the cycle it tested.

Three ways to test

OptionHow it worksSuitsWatch for
Self-contained BI with an in-clinic incubatorYou process the vial, activate it and incubate it; a colour change shows growthMost clinics testing weekly in-houseThe incubator must hold the temperature the BI’s instructions state; final read times vary by product, often a day or two
Rapid-readout BI with an auto-readerThe reader detects early signs of growth by fluorescence. Solventum’s Attest Auto-reader 490, for example, incubates its Attest steam BIs at 60 °C for a final result in 24 minutes, with software version 4.2.7 or laterBusy clinics and hospitals that want an answer the same session, including after a repairThe reader works only with its own brand’s indicators, and each test usually costs more
Spore strips sent to a laboratoryYou process strips in your steriliser and post them; the laboratory transfers them to growth medium, incubates them and reports backLow-volume users with no incubatorDays of delay before you know a cycle failed; check that a laboratory offering this service is available to you

Choosing an incubator or reader

  • Temperature: steam BIs are commonly incubated at about 55–60 °C, but the BI’s own instructions decide. Indicators for other processes, such as dry heat or ethylene oxide, use a different organism and a lower temperature.
  • Compatibility: the BI’s instructions list the incubators or readers it has been validated with. Rapid readers only read their maker’s indicators.
  • Wells: enough for the weekly test and control on every steriliser, plus repeat tests after a failure or repair.
  • Read-out: colour change read by eye, or an automatic result with a printout, data export or network link. The Attest 490, for example, comes with an ethernet cable for connection to instrument-tracking systems.
  • Temperature checks: a display or alarm, and a way to verify the block temperature with a calibrated thermometer.
  • Power: ask for the Malaysian power supply and plug version. Manufacturer listings often show a US or European plug by default.
  • Registration: in the United States, biological indicators are regulated as Class II medical devices. Ask your supplier for the MDA registration status of both the indicators and the reader.

Cost per test, and where lock-in hides

The incubator is a one-off. The indicators are the running cost, so price a year of testing, not the box.

  • Buy quantities you will use before the indicators expire, and store them as the instructions say.
  • With a rapid reader, you are committed to that brand’s indicators for the reader’s life. Ask how the indicators are supplied in Malaysia, the lead time, and whether the price is fixed for a period.
  • Ask for the BI price including controls. A low per-vial price with a separate control pack can cost more in the end.
Count thisExample: one steriliser tested weekly
Test indicators52 a year
Control indicators (one with each test)52 a year
Repeat tests after a positive result or a repairAllow a few extra
Commissioning tests for a new or moved steriliserAs your validation plan sets
Reader or incubator: purchase, service and temperature verificationOnce, then yearly

How often to test, and what a positive means

  • CDC recommends monitoring dental sterilisers at least weekly with a biological indicator and a matching control, and gives no separate schedule for part-time sterilisers.
  • CDC’s dental guidelines also recommend a BI with every load that contains an implantable device, with the result checked before the implant is used whenever possible.
  • Most policies add a BI after major repairs and at commissioning.
  • After a positive test, CDC advises repeating the test straight away on the same cycle, and recalling and reprocessing items from loads back to the last negative test where possible.

Records that make the test worth doing

  • Date, steriliser, cycle number and programme.
  • BI brand, lot number and expiry date.
  • Test result, control result, read time and who read it.
  • The cycle printout or data record for the tested load.
  • Action taken after any positive result, and the repeat-test result.

Frequently asked questions

Do I need an incubator for autoclave spore testing?
Either an in-clinic incubator for self-contained indicators, or a laboratory that receives spore strips by post and reports back. In-house testing gives you the result much sooner.
What temperature should a spore test incubator be?
The temperature in the indicator’s instructions. Steam indicators using Geobacillus stearothermophilus are commonly incubated at about 55–60 °C; the Attest Auto-reader 490, for example, runs at 60 °C.
Can I use any brand of biological indicator in my incubator?
Only indicators whose instructions list your incubator, or its temperature range for a simple dry-block unit. Rapid-readout auto-readers read only their own brand’s indicators.
How quickly do I get a spore test result?
From 24 minutes with a rapid-readout system, to a day or two for a colour-change indicator read by eye, to several days for strips posted to a laboratory.
What is in an autoclave spore test kit?
Usually test and control indicators, sometimes a test pack or process challenge device, plus an incubator and activator if you are starting from nothing, and log sheets for the results.
Can AA Alive quote a spore test incubator?
WhatsApp us your steriliser brand and model, how many sterilisers you test and how often, and we will tell you what we can quote and which indicators each option uses.

Get a written quotation