Autoclave Maintenance: Daily, Weekly and Annual Checklist
Most autoclave failures start with water, seals or skipped tests. This checklist keeps a clinic or lab steriliser passing — and lasting.
Maintenance schedule
| When | Task | Why |
|---|---|---|
| Daily | Wipe the door seal and chamber; check the water reservoir level and quality | Debris on the seal causes leaks; poor water causes scale and wet packs |
| Daily | Run the Bowie-Dick or helix test (vacuum units) | Proves air removal works before patient loads |
| Weekly | Clean trays and the chamber with the maker’s cleaner; drain and rinse the reservoir | Removes residue and mineral build-up |
| Weekly | Biological (spore) test | Definitive proof that the cycle kills spores |
| Monthly | Check and clean filters (bacteria filter, water filter, dust filter) | Blocked filters slow drying and vacuum |
| Yearly | Professional service: calibration, valves, seal replacement, safety checks | Keeps temperature and pressure readings accurate |
Read next Step-by-step cleaning scheduleSpare parts and replacement intervals
Water: the most common cause of trouble
- Use distilled or deionised water of the quality the manufacturer specifies.
- Never top up with tap or mineral water — calcium and chlorides scale valves and stain instruments.
- Empty and rinse the reservoir regularly so it never grows algae or biofilm.
- If your unit has a water-quality sensor, never bypass its warning.
Common faults and what they usually mean
| Symptom | Likely cause |
|---|---|
| Wet packs at the end of the cycle | Overloading, poor water, blocked filter, failing vacuum pump |
| Vacuum or leak test fails | Dirty or worn door seal, valve fault |
| Bowie-Dick shows uneven colour change | Air leak or poor air removal — stop using the machine |
| Stained or spotted instruments | Water quality, detergent residue, mixed metals |
| Cycle aborts on temperature or pressure | Sensor drift, heating element or valve fault — needs a technician |
Read next Troubleshooting each fault in depth
Keeping records
Keep cycle printouts or logs, daily test results, spore test results and service reports together. They are the first thing an infection-control audit or licensing inspection will ask for.
Read next Records and traceability in detailWriting an autoclave SOPServis autoclave (Bahasa Malaysia)What a service contract should cover
Related guides
Autoclave supply near you
Written quotes for clinics across Malaysia, handled from Batu Caves, Kota Kinabalu and Kuching.
Frequently asked questions
How often should an autoclave be serviced?
How often should the door seal be replaced?
Why are my autoclave packs wet?
Get a written quotation
Tell us your clinic or lab type, daily workload and location. We match suitable models and send a written quotation that states who handles delivery, installation and training, and what each costs.
010-905 2891 · Mon–Sat, 9:00 am – 6:00 pm