Skip to main content

AA Alive Sdn Bhd · SSM 1204108-DEst. 2016Superbrands Malaysia 2025Batu Caves · Kota Kinabalu · Kuching

WhatsApp 010-905 2891

Class N and Class S Autoclaves: What They Can and Cannot Sterilise

N and S are the two EN 13060 classes below Class B. They cost less and often run faster — and they are proven for a narrower set of loads. Here is where the line falls.

Also called Class N autoclave, Class S autoclave, N type, S type

Published:

Last updated:

N, S and B side by side

EN 13060, the European standard for small steam sterilisers of up to 60 litres, does not grade machines from good to bad. It classifies cycles by the loads they have been proven to sterilise. A machine is described by the cycles it offers, which is why some Class B units also carry N or S cycles.

QuestionClass NClass SClass B
Loads provenUnwrapped solid instruments onlyUnwrapped solid instruments plus the load types the maker namesWrapped and unwrapped solid, porous and hollow loads
How air is removedPassively — steam pushes the air outVaries by model: pressure pulses or a vacuum stageRepeated vacuum pulses (fractionated pre-vacuum)
Pouched instruments for storageNoOnly if wrapped loads are on the maker’s listYes
Handpieces, cannulae and other lumensNoOnly if that hollow load type is listedYes
Gauze, drapes and gownsNoOnly if porous loads are listedYes
Built-in air-removal testsNoSome modelsYes — Bowie-Dick or helix, and a vacuum leak test

What Class N is good for — and where it stops

N is often remembered as “naked”: the load goes in unwrapped and comes out for immediate use. Without a vacuum stage, steam has to push the air out of the chamber by itself. That works on smooth, open metal surfaces. It does not reliably work inside a pouch, a fold of fabric or a narrow tube, where air stays trapped and steam cannot reach the surface.

  • Suited: solid, unwrapped metal instruments — forceps, scissors, nippers, tweezers — used straight after the cycle.
  • Not suited: anything you pouch or wrap to store, because the inside of the pack cannot be shown to be sterile.
  • Not suited: hollow instruments such as dental handpieces, cannulae, suction tips or anything with a lumen.
  • Not suited: porous loads such as gauze, drapes and gowns.
  • Once the door opens, unwrapped instruments are exposed to the room. They stay sterile only until they are handled or set down, so use them at once.
  • Cycles are usually shorter, with no vacuum stage before the hold — but drying is simpler too, so check how dry a full load comes out.

Class S: read the load list, not the letter

Class S means “as specified by the manufacturer”. The maker must name the load types each S cycle has been proven on — for example unwrapped solids plus single-wrapped instruments, or certain hollow items — and the cycle is proven only for those. Two Class S machines can therefore cover very different loads.

Ask for the list in writing

The instructions for use should state the load types for each S cycle. If you are told a Class S unit “can do pouches”, ask to see where the manufacturer says so.

Check the hollow-load wording

EN 13060 separates narrow-lumen items, such as dental handpieces, from simple hollow items with wider openings. A cycle proven for one is not automatically proven for the other.

N and S cycles on a Class B machine

Some Class B units also offer N or S cycles as shorter options. Euronda’s E10 brochure, for example, lists N and S cycles that can be unlocked alongside its five B cycles where national rules allow. A fast N cycle on a B machine is still an N cycle: use it only for loads N is proven for.

Choosing between N, S and B

Choose from the most demanding load you will process, not the most common one. One pouched or hollow item a week decides the class.

Your situationUsually enoughWhy
Unwrapped solid tools, used straight after the cycleClass NMatches what N is proven for
Mostly solid tools, some single-wrapped, no lumensClass S — if wrapped loads are listedDepends entirely on the maker’s load list
Instruments pouched and stored for laterClass BOnly B is proven for all wrapped loads
Dental handpieces, cannulae, suction tipsClass BNarrow lumens need fractionated vacuum
Gauze, drapes, gowns or textile packsClass BPorous loads trap air
Not sure what you will process in two yearsClass BCovers every load type; N may limit you later

Monitoring a Class N or S autoclave

  • Record every cycle — printout, memory or a written log of time, temperature and pressure.
  • Put a chemical indicator in every load; for unwrapped instruments, place it on the tray among them.
  • Bowie-Dick and helix tests are designed for vacuum cycles. A Class N unit without a vacuum stage does not run them; a Class S unit with one may — follow the manufacturer.
  • Run biological (spore) tests at the interval your infection-control policy sets, and after any repair.
  • Give it the same cleaning, water and door-seal care as any autoclave: dirty water and a worn seal fail N cycles just as they fail B cycles.

Class N autoclaves we supply, with prices

Three Tonsor Class N models for unwrapped solid instruments. If anything you sterilise is pouched or hollow, choose Class B instead. Prices are for one unit; delivery, installation and warranty are listed on the written quotation.

  • Tonsor T&S series tabletop autoclave (17 L pictured; the 23 L is longer)

    Class N autoclave, 23 L

    Tonsor 23 L Class N

    Practices that sterilise only unwrapped solid instruments, such as beauty, nail and tattoo studios.

    • Class N steam sterilisation (no vacuum), for unwrapped solid instruments
    • Chamber 23 L, ø25 × 49 cm, 2 trays
    • Programs at 134°C and 121°C; LCD display
    • 230 V, 1500 W; 36.5 kg

    RM 5,400Price for one unit, supplied on indent order. Delivery, installation and warranty are listed line by line on the written quotation. The autoclaves we supply are MDA-registered, and we handle the JKKP (DOSH) approval and renewal for them.

  • Tonsor T&S series tabletop autoclave (17 L pictured; the 29 L is longer)

    Class N autoclave, 29 L

    Tonsor 29 L Class N

    Higher volumes of unwrapped solid instruments where no vacuum cycle is needed.

    • Class N steam sterilisation (no vacuum), for unwrapped solid instruments
    • Chamber 29 L, ø25 × 59 cm, 2 trays
    • Programs at 134°C and 121°C; LCD display
    • 230 V; 80 × 48 × 42 cm

    RM 8,700Price for one unit, supplied on indent order. Delivery, installation and warranty are listed line by line on the written quotation. The autoclaves we supply are MDA-registered, and we handle the JKKP (DOSH) approval and renewal for them.

  • Tonsor 45 L series autoclave (Class B unit pictured)

    Class N autoclave, 45 L

    Tonsor 45 L Class N

    Large batches of unwrapped solid instruments.

    • Class N steam sterilisation (no vacuum), for unwrapped solid instruments
    • Chamber 45 L, ø31.6 × 62 cm, 2 trays
    • Programs at 134°C and 121°C; LCD display with keypad
    • 220 V; 83 × 65 × 55 cm

    RM 16,400Price for one unit, supplied on indent order. Delivery, installation and warranty are listed line by line on the written quotation. The autoclaves we supply are MDA-registered, and we handle the JKKP (DOSH) approval and renewal for them.

Frequently asked questions

What does Class N mean on an autoclave?
It is the EN 13060 class for cycles proven only on unwrapped solid instruments. Air is removed passively, without a vacuum pump, so wrapped, porous and hollow loads are not covered.
Can a Class N autoclave sterilise instruments in pouches?
No. Without a vacuum stage, air can stay trapped inside the pouch, so the contents cannot be relied on as sterile. Use a Class B cycle, or a Class S cycle whose load list includes wrapped instruments.
Is Class S as good as Class B?
Only for the loads the manufacturer lists for that S cycle. Class B is proven for all the load types in the standard; Class S is proven for a named subset.
Is a Class N autoclave enough for a tattoo or nail studio?
For unwrapped solid metal tools used straight after the cycle, it can be. If you pouch tools to keep them sterile until each client, or use anything hollow, choose Class B.

Get a Class N quotation

Tell us what tools you sterilise and whether any are pouched. We confirm Class N is enough and send the price in writing.

010-905 2891 · Mon–Sat, 9:00 am – 6:00 pm